5 year Radiostereommetry, Bone Mass Density,

and Whole Blood Cr Co levels in Resurfacing (RHA),

Large Diameter Head (LDH) and Standard THA.

Results from Randomized studies

Jeannette Penny, Ming Ding, Ole Ovesen, Søren Overgaard

Dept. of orthopedics, Næstved and OUH; , Clinical Institute OUH; Dept. of or-

thopedics, OUH; Dept. of orthopedics, Clinical Institute OUH

Background:

Most Metal-on-metal hips were abandoned due to high clinical

failure rates likely caused by metal wear products.

Purpose / Aim of Study:

We report 5 year results of a randomized study of

RHA, LDH-THA and THA with implant micro motion, Cr & Co ions and BMD.

Links between outcome measures is investigated

Materials and Methods:

19 RHA 17 THA and 15 LDH-THA for 5 year follow-

up All were followed with Bone mass density RHA and THA had prospective

Co and Co ions, LDH-THA only at 5 y. Marker based RSA of both RHA com-

ponents. Cup only for LDH-THA. We used a total translation measure: TT= ã

(x2+y2+z2). Data collected at baseline, 8 w, 6 m 1, 2 and 5 y Statistical tests:

Pearson correlations. ANOVA and MANOVA

Findings / Results:

Presented as: median (mean (s.e)) and quartiles RSA: The

5 y TT results were 1.1(1.3 (0.2))1.0 to 1.7 mm for the LDH-THA Cup, 0.6

(0.7(0.1)) 0.3 to 0.8 mm for the RHA Cup and 0.5 (0.8(0.3)) 0.3 to 0.6 mm

for the RHA head. The migration between the cups differed at 5 but not at 2

years (p<0.05) Co ions were; RHA: 1.1(2.8 (0.7)) 0.8 to 4.9, THA: 0.2 (0.4

(0.1)) 0.1 to 0.4 and LDH-THA: 1.2 (1.4(0.3)) 0.7 to 2.9 ppb. Similar levels

for Cr. LDH-THA and RHA ions did not differ at 5 year, Co was higher than THA

(p<0.01) We found better preservation of the acetabular BMD in the LDH-RHA

group (p<0.05) and an advantage of the RHA in the femoral Gruen zones 6 and

7 (p<0.01) TT migration correlated negatively with BMD in the femoral collum

and proximal femur (p<0.05), but had no link to acetabular BMD or metal ions

Conclusions:

In contrast to registered revision rates, we found significantly

larger movement for the LDH-THA cup than the RHA cup. Despite this, the

LDH-THA cup maintained the acetabular BMD best at 5 years. Movement of

the implant may adversely affect the femoral BMD, but - in this study - does

not increase metal ion release

No conflicts of interest reported