DOS Kongressen 2017 ·

127

POLYETHYLENE WEAR OF A DUAL-MOBILITY ARTICU-

LATION IN TOTAL TRAPEZIOMETACARPAL ARTHRO-

PLASTY

Maiken Stilling, Lene Dremstrup, Lone Kirkeby, Torben Bæk Hansen

Department of Orthopaedics, University Clinic for Hand, Hip and Knee Surgery,

Holstebro;

Background:

Dual-mobility (DM) articulation provides good stability in tra-

peziometacarpal (TMC) arthroplasty but the thin polyethylene (PE) liner raises

concern about possible wear issues.

Purpose / Aim of Study:

To evaluate cup fixation and PE wear of a press-fit

conical DM cup.

Materials and Methods:

A case-study of the first 111 consecutive patients

(111 hands, 89 females) with Eaton stage 2-4 osteoarthritis (June 2013-May

2015). Moovis Elektra 9mm conical press-fit cup with UHMWPE DM liner and

5mm CoCr metacarpal head was used. Model-based radiostereometry (RSA)

was used to measure cup fixation and PE wear. Cup subsidence was evaluated

in the length axis of the cup. Polyethylene wear was measured as head/neck

migration with respect to cup model feature points. Double RSA examinations

were obtained for precision (95% agreement limits). RSA and PROMs were

evaluated at baseline, 3 months, 1 and 2 years.

Findings / Results:

Patients were mean 58 years (range 42-76). Precision

was 0.15 mm. Cup subsidence was 0.02 mm (SD 0.35) at 2 years. Total cup

translation (TT) at 2 years was mean 0.46 (SD 0.48) mm, and increased from

3 months to 2 years (p=0.01). At 2 years 13 cups (15%) had migrated more

than 1mm TT (p=0.00), but no cups were revised. 3 trapeziums fractured intra-

operatively and 2 patients had intraprosthetic dislocation of the dual-mobility

liner. Clinical function improved (p=0.000) by VAS_rest 3 (SD 3), VAS_activ-

ity 5 (SD3), and qDASH 31 (SD 20). Bedding-in of the PE liner was 0.15mm

(SD 0.16) until 3 months (p=0.001). PE wear was 0.06mm (0.17mm) from

3 months until 2 years (p=0.002), and 0.003mm (SD 0.13) between 1 and 2

years.

Conclusions:

At 2 years followup Moovis Elektra cup fixation was acceptable

for most implants (85%) and clinical improvements significant. There was bed-

ding-in of the PE liner until 3 months, but after 1 year PE wear was small.

No conflicts of interest reported