Previous Page  197 / 245 Next Page
Information
Show Menu
Previous Page 197 / 245 Next Page
Page Background

DOS Kongressen 2016 ·

197

Agreement and reliability of acetabular Bone

Mineral Density measurements in total hip

arthroplasty using Single and Dual energy

computed tomography with 3-dimensional segmentation

Bo Mussmann, Søren Overgaard, Trine Torfing, Morten Bøgehøj, Oke Gerke, Poul

Erik Andersen

Department of Radiology, Odense University Hospital; Department of Orthopedic

Surgery and Traumatology, Odense University Hospital; Department of Radiology,

Odense University Hospital; Department of Orthopedic Surgery and Traumatology,

Odense University Hospital; Department of Nuclear Medicine, Odense University

Hospital; Department of Radiology, Odense University Hospital

Background:

Periprosthetic bone loss is considered a predictor of aseptic loosening of

the acetabular component in total hip arthroplasty. However, no studies have shown this

association. This may be explained by imaging methods used. Dual energy CT (DECT) has

previously shown better delineation of the interface between bone and prosthesis and

may be beneficial in quantitative analysis of bone loss close to the implant as compared

to single energy computed tomography (SECT).

Purpose / Aim of Study:

To test the agreement and reliability of bone mineral density

measurements (BMD) in close proximity of the acetabular cup using SECT and DECT im-

ages and 3D segmentation software.

Materials and Methods:

12 un-cemented and 12 cemented cups were inserted in por-

cine hip specimens ex vivo. A femoral stem was attached to each specimen and imag-

ing was performed with SECT and DECT. The specimens were repositioned and scans

repeated to obtain double measurements. For each scan BMD was measured in a hemi-

spherical volume around the acetabular cup using in-house segmentation software.

Findings / Results:

In the uncemented concept mean BMD difference between the dou-

ble measurements in SECT was 8 mg/cm3 (p=0.64) and 2 mg/cm3 in DECT (p=0.596).

ICC was 0.90 for SECT and 0.91 for DECT. In the cemented concept the differences were

41 mg/cm3 (p=0.055) and 11 mg/cm3 (p=0.013), respectively, and ICC was 0.74 for

SECT and 0.91 for DECT. In both concepts the Bland Altman limits of agreement were

wider in SECT (uncemented: -95 to 111; cemented: -107 to 189) compared with DECT

(uncemented: -28 to 23; cemented: -20 to 42).

Conclusions:

There were no statistically significant reliability differences between SECT

and DECT, but results suggest that the agreement of DECT is better than SECT, and both

scan modes perform better in the un-cemented concept compared with the cemented

concept.

No conflicts of interest reported

148.